Related Experiment Video
Updated: Jul 5, 2026

Transthoracic Speckle Tracking Echocardiography for the Quantitative Assessment of Left Ventricular Myocardial Deformation
Published on: October 20, 2016
Quantitative assessment of left ventricular volume and ejection fraction using two-dimensional speckle tracking
Tomoko Nishikage1, Hiromi Nakai, Victor Mor-Avi
1Echocardiographic Laboratory, Tane General Hospital, Osaka, Japan.
Aims:
Two-dimensional speckle tracking echocardiography (2DSTE) allows measurements of left ventricular (LV) volumes and LV ejection fraction (LVEF) without manual tracings. Our goal was to determine the accuracy of 2DSTE against real-time 3D echocardiography (RT3DE) and against cardiac magnetic resonance (CMR) imaging.
Methods And Results:
In Protocol 1, 2DSTE data in the apical four-chamber view (iE33, Philips) and CMR images (Philips 1.5T scanner) were obtained in 20 patients. The 2DSTE data were analysed using custom software, which automatically performed speckle tracking analysis throughout the cardiac cycle. LV volume curves were generated using the single-plane Simpson's formula, from which end-diastolic volume (LVEDV), end-systolic volume (LVESV), and LVEF were calculated. In Protocol 2, the 2DSTE and RT3DE data were acquired in 181 subjects. RT3DE data sets were acquired, and LV volumes and LVEF were measured using QLab software (Philips). In Protocol 1, excellent correlations were noted between the methods for LVEDV (r=0.95), ESV (r=0.95), and LVEF (r=0.88). In Protocol 2, LV volume waveforms suitable for analysis were obtained from 2DSTE images in all subjects. The time required for analysis was <2 min per patient. Excellent correlations were noted between the methods for LVEDV (r=0.95), ESV (r=0.97), and LVEF (r=0.92). However, 2DSTE significantly underestimated LVEDV, resulting in a mean of 8% underestimation in LVEF. Intra- and inter-observer variabilities of 2DSTE were 7 and 9% in LV volume and 6 and 8% in LVEF, respectively.
Conclusions:
Two-dimensional speckle tracking echocardiography measurements resulted in a small but significant underestimation of LVEDV and EF compared with RT3DE. However, the accuracy, low intra- and inter-observer variabilities and speed of analysis make 2DSTE a potentially useful modality for LV functional assessment in the routine clinical setting.
Related Concept Videos
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...