First day of life pulse oximetry screening to detect congenital heart defects
Alf Meberg1, Sabine Brügmann-Pieper, Reidar Due
1Department of Paediatrics, Vestfold Hospital, Tønsberg, Norway. alfmeb@start.no
Objective:
To evaluate the efficacy of first day of life pulse oximetry screening to detect congenital heart defects (CHDs).
Study Design:
We performed a population-based prospective multicenter study of postductal (foot) arterial oxygen saturation (SpO(2)) in apparently healthy newborns after transfer from the delivery suite to the nursery. SpO(2) < 95% led to further diagnostic evaluations. Of 57,959 live births, 50,008 (86%) were screened. In the screened population, 35 CHDs were [corrected] classified as critical (ductus dependent, cyanotic). CHDs were prospectively registered and diagnosed in 658/57,959 (1.1%) [corrected]
Results:
Of the infants screened, 324 (0.6%) failed the test. Of these, 43 (13%) had CHDs (27 critical), and 134 (41%) had pulmonary diseases or other disorders. The remaining 147 infants (45%) were healthy with transitional circulation. The median age for babies with CHDs at failing the test was 6 hours (range, 1-21 hours). For identifying critical CHDs, the pulse oximetry screening had a sensitivity rate of 77.1% (95% CI, 59.4-89.0), specificity rate of 99.4% (95% CI, 99.3-99.5), and a false-positive rate of 0.6% (95% CI, 0.5-0.7).
Conclusions:
Early pulse oximetry screening promotes early detection of critical CHDs and other potentially severe diseases. The sensitivity rate for detecting critical CHDs is high, and the false-positive rate is low.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Fetal Circulation
Two umbilical arteries transport blood from the fetus to the placenta. At the placenta, the blood absorbs oxygen and nutrients while simultaneously eliminating waste products. This oxygen-enriched and nutrient-rich blood then returns to the fetus through one...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for: