Related Experiment Video
Updated: Jul 5, 2026

Bimolecular Fluorescence Complementation
Published on: April 15, 2011
Combined fluorescent and interferometric detection of protein on a BioCD
Xuefeng Wang1, Ming Zhao, D D Nolte
1Department of Physics, Purdue University, West Lafayette, IN 47907-2036, USA.
Abstract:
We perform simultaneous interferometric and fluorescent detection of molecular protein layers on a BioCD. The 488 nm excitation wavelength of fluorescein also provides the interferometric detection channel that operates in a common-path in-line configuration in the condition of phase quadrature set by a thermal oxide on silicon. The simultaneous acquisition of both channels enables a direct correlation between bound mass and fluorescent surface density, which we compare in forward- and reverse-phase immunoassays. Scaling mass sensitivities for immunoassays measured in the interferometric and fluorescent channels are 15 pg/mm and 1.5 pg/mm, respectively, when applied to gel-printed periodic antibody patterns detected in the frequency domain from the spinning disc. These sensitivities are limited by the inhomogeneities of the print. While fluorescence is subject to bleaching, the interferometry signal is robust under long-term laser illumination.
Related Concept Videos
Protein Dynamics in Living Cells
Fluorescent recovery after photobleaching (FRAP) is a fluorescent-protein-based detection technique used to quantify protein movement rates within the cell. This method exposes a small portion of the cell to an intense laser beam. The laser beam causes permanent photobleaching of the fluorophore-tagged proteins in the exposed region. As the bleached...