Related Experiment Video
Updated: Jul 5, 2026

Optical Screening of Novel Bacteria-specific Probes on Ex Vivo Human Lung Tissue by Confocal Laser Endomicroscopy
Published on: November 29, 2017
Autofluorescence Detection of Bronchial Tumors With the D-Light/AF
K Häußinger1, F Stanzel, R M Huber
1Clinic for Pneumology Fachklinik München - Gauting Robert-Koch-Allee 2 Gauting D-82131 Germany.
Abstract:
We present a newly developed diagnostic system combining a conventional light source (white light mode and two different fluorescence excitation modes), a bronchoscope and optionally a highly sensitive camera (Baumgartner et al., Photochem. Photobiol. 1987; 46(5): 759-763). Routine diagnostics can be performed with the autofluorescence bronchoscopy (AFB) and the white light bronchoscopy (WLB) in one diagnostic procedure. The image is visible directly with the naked eye. The system was evaluated in a pilot study including 60 patients. Two hundred and sixty-four biopsies were taken to detect premalignant and malignant findings (Stanzel et al., Contribution to 10th World Congress for Bronchology, June 1998). The sensitivity of the combination of WLB and AFB was 2.8 times higher than that of the conventional WLB. The specificity decreased from 94% (WLB) to 89% (WLB + AFB). The results of this preliminary pilot study are being confirmed in a multicenter study, which will begin at seven European centers.