A new method for the determination of aortic pulse wave velocity using cross-correlation on 2D PCMR velocity data
Samuel W Fielden1, Brandon K Fornwalt, Michael Jerosch-Herold
1Department of Radiology, Emory University School of Medicine, Atlanta, Georgia 30322, USA.
Purpose:
To evaluate the reproducibility of a new multisite axial pulse wave velocity (PWV) measurement technique that makes use of 2D PCMR data and cross-correlation analysis.
Materials And Methods:
PWV was estimated with MRI in 13 healthy volunteers by a transit-time technique (TT), a multisite technique utilizing 1D PCMR data in the descending aorta (FOOT), and a new multisite axial technique that uses 2D PCMR data over the ascending, transverse, and descending sections of the aorta (2D-XC).
Results:
No significant difference was observed between PWV measurements values measured by the three techniques. However, 2D-XC displayed significantly better intertest reproducibility than either the TT or FOOT methodologies. Average percent difference between scans: TT: 15.8% +/- 13.4%, FOOT: 21.3% +/- 16.9%, 2D-XC: 7.72% +/- 4.73%. P = 0.02 for both 2D-XC/TT comparison and 2D-XC/FOOT comparison.
Conclusion:
2D-XC is a more reproducible method than either the established TT or FOOT methods to estimate the aortic PWV.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(two-step method)
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Pulse
The pulse serves as a clinical indicator...
Special considerations while measuring pulse