Effects of scatter radiation on ICD and CRT function
Suraj Kapa1, Luis Fong, Charles R Blackwell
1Department of Internal Medicine, Mayo Clinic, Rochester, Minnesota 55905, USA.
Scatter radiation from radiotherapy did not cause malfunctions in implantable cardioverter-defibrillators (ICDs) or cardiac resynchronization therapy (CRT)-ICDs in this study. While rare, precautions are advised during radiation therapy for patients with these cardiac devices.
Area of Science:
- Cardiology
- Radiation Oncology
- Biomedical Engineering
Background:
- Direct radiation effects on cardiac devices are known, but scatter radiation impacts remain unclear.
- Existing recommendations for radiotherapy in patients with cardiac devices are based on limited pacemaker studies.
- The effects of scatter radiation on implantable cardioverter-defibrillators (ICDs) and cardiac resynchronization therapy (CRT)-ICDs require further elucidation.
Purpose of the Study:
- To investigate the impact of scatter radiation on ICDs and CRT-ICDs.
- To assess device function and programmed parameters following exposure to scatter radiation.
Main Methods:
- Exposed 12 ICDs and 8 CRT-ICDs to 400 cGy scatter radiation from a 6-MV photon beam.
- Performed device interrogations before, during, and after radiation exposure, plus a 1-week follow-up.
- Retrospectively reviewed 13 patients with cardiac devices undergoing radiotherapy (device outside radiation field), with 12 interrogated pre- and post-radiation.
Main Results:
- No instances of device reset or malfunction were observed during or after scatter radiation exposure.
- Patient device interrogations revealed no episodes of reset, inappropriate sensing, or therapy changes.
- Programmed parameters remained stable throughout the study period for all tested devices and patients.
Conclusions:
- Scatter radiation is unlikely to cause device reset or malfunction in ICDs and CRT-ICDs.
- No clear contraindications exist for radiotherapy in patients with ICDs or CRT-ICDs.
- Precautions to avoid direct radiation and close patient monitoring before and after radiotherapy are recommended.
Related Concept Videos
Biological Effects of Radiation
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...