Related Experiment Video
Updated: Jul 4, 2026

Extraction and Detection of Geosmin and 2-Methylisoborneol in Water and Fish using High-Capacity Sorptive Extraction Probes and GC-MS
Published on: July 3, 2025
Determination of mercury in water and fish samples by cold vapor atomic absorption spectrometry after solid phase
1Department of Chemistry, College of Science, Shahid Chamran University, Ahvaz, Iran. npourreza@yahoo.com
Abstract:
A novel solid phase extraction method for the determination of mercury has been developed. The Hg(II) ions were retained on a mini-column packed with agar powder modified with 2-mercaptobenzimidazole at a flow rate of 6 mL min(-1). The retained Hg(II) ions were eluted with 3 mol L(-1) solution of HCl and measured by cold vapor atomic absorption spectrometry (CV-AAS). The effect of different variables such as pH, sample flow rate, amounts of 2-mercaptobenzimidazole loaded on agar and SnCl(2) concentration was investigated and optimum conditions were established. The calibration curve was linear in the range of 0.040-2.40 ng mL(-1) with r=0.9994 (n=8). The limit of detection based on three times the standard deviation (3S(b)) (n=10) obtained under optimum conditions was 0.02 ng mL(-1). The relative standard deviation (R.S.D.) for the determination of 0.4 and 2.0 ng mL(-1) of Hg(II) was 2.6 and 1.9% (n=8), respectively. The method was successfully applied to determine Hg(II) in water, wastewater and fish samples.