Indirect measurement of systemic blood pressure in conscious dogs in a clinical setting
Tsai-Yuan Hsiang1, Yu-Hsin Lien, Hui-Pi Huang
1Department of Veterinary Medicine, National Taiwan University, National Taiwan University Hospital, Taipei, Taiwan.
Abstract:
The present study compared 2 indirect methods, Doppler sphygmomanometry and oscillometry, for measurement of the systemic blood pressure level in 100 conscious, client-owned dogs in a clinical setting on 2 separate occasions. The mean systemic blood pressure values, measured by Doppler sphygmomanometry on 2 separate occasions, were 156 +/- 38.2 mmHg and 150 +/- 34.1 mmHg, respectively. Using oscillometry, the mean systolic blood pressure values were 138 +/- 36.9 mm Hg and 133 +/- 33.5 mm Hg on 2 separate occasions. There were significant differences between the systemic blood pressure readings for both methods on the same occasion (P<0.001). The coefficients of variance from 5 consecutive measurements in the same dog obtained by Doppler sphygmomanometry on the 2 separate occasions were 4.1 +/- 3.2% and 3.1 +/- 1.7%; that of the oscillometric method on the 2 separate occasions were 18.7 +/- 11.3% and 17.2 +/- 12.5%. The coefficients of variance of these 2 methods were statistically different on each occasion (P<0.001). Five consecutive systemic blood pressure readings were obtained for each dog within 6 min on both occasions using Doppler sphygmomanometry. More than 15 min was required to complete 5 consecutive systemic blood pressure readings by oscillometric sphygmomanometry for all dogs on each occasion. The results of this study indicate that Doppler sphygmomanometry provides more efficient and precise measurements of the systemic blood pressure level than oscillometric testing in conscious dogs in a clinical setting.
Related Concept Videos
Measurement of Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...