Abstract:
The purpose of this study was to evaluate the rotational kinematics of a fixed-bearing posteriorly stabilized total knee design in moderate and deep flexion. Three-dimensional kinematics analyses were conducted on 20 knees in 4 weight-bearing positions using 3-dimensional shape-matching techniques. Average maximum skeletal flexion was 138 degrees . Internal tibial rotation was demonstrated in 19 of 20 knees. The average internal tibial rotation in midflexed lunge was 5.5 degrees (-3.8 degrees to 14.1 degrees ) and in maximum flexion kneeling was 4.0 degrees (-3.1 degrees to 10.6 degrees ). Separation of articular surfaces was not identified. In this study, patients with this device demonstrated patterns of rotation similar to those previously reported for both the normal knee and rotating platform designs.