Measurement of oxygen saturation in the retina with a spectroscopic sensitive multi aperture camera
Jessica C Ramella-Roman1, Scott A Mathews, Haryipriya Kandimalla
1The Catholic University of America, Washington, D.C., USA. ramella@cua.edu
Abstract:
We introduce a new multi aperture system capable of capturing six identical images of the human fundus at six different spectral bands. The system is based on a lenslet array architecture and is well suited for spectroscopy application. The multi-aperture system was interfaced with a fundus camera to acquire spectroscopic sensitive images of the retina vessel and ultimately to calculate oxygen saturation in the retina in vivo. In vitro testing showed that the system is able to accurately reconstruct curves of partially oxygenated hemoglobin. In vivo testing on healthy volunteers was conducted and yielded results of oxygen saturation similar to the one reported in the literature, with arterial SO(2) approximately 0.95 and venous SO(2) approximately 0.5.
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.