Related Experiment Video
Updated: Jul 4, 2026

Microalgae Cultivation and Biomass Quantification in a Bench-Scale Photobioreactor with Corrosive Flue Gases
Published on: December 19, 2019
Design of laboratory continuous-culture equipment for accurate gaseous metabolism measurements
J D Brooks1, D G Maclennan, J P Barford
1Department of Chemical Engineering, University of Sydney, New South Wales 2006, Australia.
Abstract:
The accuracy of kinetic and stoichiometric data obtained from most laboratory-scale continuous-culture equipment, particularly involving gaseous measurements, may be much lower than many workers realize, despite the use of good quality instruments. For example, errors in specific oxygen uptake measurements (QO(2)) easily can be as high as +/-100%. This article assesses the accuracies of individual instruments and of the overall system in greater detail than has previously been reported and suggestions are made as to how the errors can be reduced to acceptable levels.
Related Concept Videos
Bioreactor Design and Operational System
Designing Growth Media for Bioreactors
Bioreactor Controls-I
Bioreactor Controls-II
Scale-Up Processes