Development of a device to standardize leak point pressure experiments in rats
Andrew J Shoffstall1, Paul J Zaszczurynski, Robert S Butler
1Department of Biomedical Engineering, Cleveland Clinic, Cleveland, OH 44195, USA.
Aims:
The objective of this study was to test a custom device aimed at increasing repeatability of abdominal leak point pressure (LPP) measurements in rodents.
Methods:
The device consisted of a soft-tipped force applicator with a force sensor which was moved in the vertical direction by a linear actuator, laser crosshairs for accurate positioning, and a hand-held wired remote control system. One expert and two novice experimenters acquired LPP measurements using manual and device methods by applying a gradually increasing force with the finger or the device, respectively, until a leak was visually observed at the urethral meatus. The device was tested at fast, medium, and slow speeds. A leak sensor was also tested to determine presence of a leak and reaction time of the user. The change in bladder pressure due to the externally applied force (LPP) was the primary outcome.
Results:
There were no significant differences in mean LPP value when compared across experiments (expert, novice), method (manual, device), or speeds. The pooled variance of LPP was significantly decreased (P < 0.05) compared to the manual method when using the device at medium speed; however the slower speed showed no further improvement over the medium speed. The wet sensor detected leaks 385 +/- 187 msec earlier than users' reaction.
Conclusions:
Use of the device can decrease variability of the experiment but does not alter mean values. Reaction time, inherent in both the manual and device methods, has the potential for a significant impact on LPP outcomes.