Measurements of lactate concentration using lactate oxidase and an electrochemical oxygen sensor
1Chemical Engineering Department, Case Western Reserve University, Cleveland, Ohio 44106.
Abstract:
Lactate oxidase was used in combination with an electrochemical dissolved oxygen sensor to measure L-lactate concentration in the physiological saline solution. The rate of oxygen consumption was found to have an excellent linear relationship with the lactate concentration on a log-log scale in the lactate concentration range of 0.3-25 mM. This detection does not require additional reagents and can be developed into a simple method of L-lactate determination. The effects of the temperature and pH of the solution on the reaction rate were investigated and discussed.