Determination of urinary iodine by inductively coupled plasma mass spectrometry
P Macours1, J C Aubry, B Hauquier
1Laboratory of Clinical Chemistry, Hôpital Erasme, Université Libre de Bruxelles, Route de Lennik 808, 1070 Bruxelles, Belgium.
Background:
Mild iodine deficiency is endemic in many countries of Europe including Belgium. Fast, accurate and specific methods for quantification of urinary iodine are needed. We describe in this report a specific ICP-MS method for the quantification of urinary iodine.
Method:
Samples and iodate calibrators were diluted 20 times into aqueous solution containing triton X-100, 1.5% HCl and (103)Rh as an internal standard. Prior digestion or oxidation was not necessary. Results were compared with those obtained by Sandell-Kolthoff (S-K) spectrophotometric method.
Results:
Comparison of both methods showed good agreement. The Passing-Bablok regression between both methods was ICP-MS=0.986 (S-K)-7.51. The Bland-Altman difference plot showed a small but significant mean difference of -13.3 microg/L for ICP-MS. The between-day coefficient of variation (CV) was 13% at 89 microg/L. Limit of detection was 4 microg/L and limit of quantification was 20 microg/L. No carryover effect has been observed on series containing up to 50 samples.
Conclusion:
The ICP-MS method described here is fast, accurate and specific for the quantification of urinary iodine. Compared to the S-K method the urinary iodine concentrations measured by the ICP-MS method were slightly, but significantly lower. Consequently, the results of studies using S-K method should be compared with caution with those using the ICP-MS method.
More Related Videos
09:22Absolute Quantitation of Inositol Pyrophosphates by Capillary Electrophoresis Electrospray Ionization Mass Spectrometry
Published on: August 13, 2021
08:21Determination of Inorganic Arsenic in a Wide Range of Food Matrices using Hydride Generation - Atomic Absorption Spectrometry.
Published on: September 1, 2017
Related Concept Videos
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Redox Titration: Iodimetry and Iodometry
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Ion-Exchange Chromatography