[Biocompatibility of polyurethane-BaFe(12)O(19 ) composite microsphere as a new endovascular embolization material]
Zai-Yu Guo1, Lian-Ting Ma, Jun Li
1Department of Neurosurgery, Wuhan General Hospital of Guangzhou Military Command, Wuhan 430070, China. guozai_yu@163.com
Objective:
To investigate the biocompatibility of polyurethane-BaFe(12)O(19) magnetic composite microsphere as a new endovascular embolization material.
Methods:
The biocompatibility of BaFe(12)O(19) particle was evaluated in vitro using Ames test, cell toxicity test, acute and subacute systemic toxicity test, hemolysis test, bleeding time and clotting time test and blood clotting function assay.
Results:
Ames test showed that the MR values of this particle leaching solution were all less than 2 without mutagenicity. Cell toxicity test showed that leaching solution at different concentrations had grade I toxicity on L929 cells. Acute and subacute systemic toxicity test showed that the experimental animals had good general condition without obvious pathological abnormality. The hemolysis rate of experimental group was 2.43%, which met the ISO standard (no more than 5%). The bleeding time and clotting time in mice were comparable between the experimental group and control group (P>0.05). There were no significant differences in blood clotting function between experimental group and control group (P>0.05).
Conclusion:
The material has no obvious toxicity or mutagenicity, and does not cause hemolysis or hemopexis or affect the bleeding time and clotting time. Polyurethane-BaFe(12)O( 19) particle possesses satisfactory biocompatibility.