Analyzing the Photo-oxidation of 2-propanol at Indoor Air Level Concentrations Using Field Asymmetric Ion Mobility Spectrometry
Published on: June 14, 2018
[Study on method for benzene and its homologues in indoor air monitored by portable gas chromatography]
Chunyu Xu1, Dongqun Xu, Xiaoyan Dong
1Institute for Environmental Health and Related Products Safety, Chinese Center for Disease Control and Prevention, Beijing 100021, China.
Objective:
To set up a method for the low concentration of benzene and its homologues (toluene, ethylbenzene and xylene, BTEX) in indoor air monitored simultaneously by portable gas chromatography (GC).
Methods:
The target gas of individual benzene and its homologues was prepared in Tedlar bag by liquid standard solution. The air samples were collected with the on- board sampling pump and preconcentrated on a sorbent tube, then thermally desorbed into capillary column and detected by MAID detector to determine the retention time. Retention time was used to confirm the target compound. The mix calibration gas of benzene and its homologues was prepared by liquid external standard method. Peak area was used to determine the amount of mixture.
Results:
The benzene and its homologues were separated very well in the selected analysis parameter. The analytical period was less than 15 min. The concentration ranges that could be determined were 0.018-0.879 mg/m3 of benzene, and 0.035-1.299 mg/m3 of toluene, and 0.035-1.300 mg/m3 of ethylbenzene, and 0.035-1.320 mg/m3 of m, p-xylene, and 0.035-1.296 mg/m3 of o-xylene, respectively. The minimum detectable concentration of benzene, toluene, m, p-xylene and o-xylene were 7.5, 11.5, 28.8, 30.3 and 30.6 microg/m3 respectively. The relative deviations between this method and standard method were less than 13.9% and 20.0% in laboratory and in filed study, respectively. In the precision experiment, the relative standard deviations of BTEX were less than 8.4% in different concentrations within one day and one week.
Conclusion:
This method which has high sensitivity and precision, and strong anti-interference ability could be used to detect the low concentration of benzene and its homologues in indoor air.
Related Concept Videos
NMR Spectroscopy of Benzene Derivatives
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Aromatic Compounds: Overview
In 1825, Faraday isolated benzene...
Gas Chromatography: Types of Detectors-II
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...