Related Experiment Video
Updated: Jul 4, 2026

A Protocol for Decellularizing Mouse Cochleae for Inner Ear Tissue Engineering
Published on: January 1, 2018
[Computer-assisted manufacture of individual auricular framework and the experimental study for ear reconstruction]
Jie Luan1, Ting-Chun Shi, Ren-Ji Zhang
1Plastic Surgery Hospital, Chinese Academy of Medical Science, Beijing 100041, China.
Objective:
To manufacture the individual polyurethane (PUR) auricular framework using the rapid prototyping (RP) technique and to evaluate its feasibility in ear reconstruction.
Methods:
3-D models of the patient's auricle were reconstructed according to the computed tomography (CT) data. The supporting and drainage structures were created. Then the individual PUR auricular frameworks were manufactured using RP technique. The frameworks were tested for the fatigue strength and elasticity. The frameworks were also put at the subcutaneous layer of rat. At 1, 2, 4, 8, 12 weeks after operation, the shape of the reconstructed ears was observed and the histological examination was performed.
Results:
The PUR framework had good elasticity and a much better fatigue strength than high density polyethylene (HDPE) framework. The shape of reconstructed ear matched the prototype very well. The around tissue grew into the implant pore and adhered tightly to the framework 2 weeks after implantation. Histologic examination showed integrated capsule four weeks later without lymphocytes infiltration. The shape kept very well twelve weeks later.
Conclusions:
PUR auricular framework manufactured by the 3D reconstruction and RP techniques has very good shape, intensity and elasticity. It can be selected to replace the autograft of rib cartilage framework.