Intrapulmonary percussive ventilation superimposed on conventional ventilation: bench study of humidity and
Jean Dellamonica1, Bruno Louis, Aissam Lyazidi
1Réanimation Médicale, AP-HP, Centre Hospitalier Albert Chenevier, Henri Mondor, Créteil, France.
Objective:
Intrapulmonary percussive ventilation (IPV) is a form of high-frequency ventilation that can be superimposed on spontaneous breathing or conventional ventilation. Drawbacks include difficulties achieving adequate airway humidification and an inability to monitor delivered volumes and pressures, which may vary with patient characteristics. The objectives of this study were to assess various humidification set-ups, to measure intrapulmonary pressures and volumes resulting from IPV superimposed on a conventional driving ventilator (DV) and to test several ventilators regarding their ability to accept added IPV.
Design:
Bench study in a test-lung set-up was used to measure humidification and the effects of adding IPV to a DV under various conditions of compliance, resistance, plateau and positive end-expiratory pressures. Then, five ventilators were tested in combination with IPV.
Measurements And Results:
Adequate humidification required a heated humidifier on the inspiratory line downstream of the IPV device. IPV increased end-inspiratory intrapulmonary pressures up to 10 cmH(2)O, increased delivered volumes up to 237 ml and generated intrinsic PEEP from 1.7 to 4.3 cmH(2)O when no PEEP was set on the DV. Intrinsic PEEP was lower or absent when PEEP was set on the DV. With most tested ventilators, IPV prevented reliable flow monitoring. Autotriggering and missing cycles were common and the PEEP effect varied across DVs.
Conclusion:
Achieving adequate humidification with IPV requires a specific set-up. Superimposing IPV on standard ventilation can increase intrapulmonary pressures and tidal volumes importantly and interfere with the triggering sensors of the ventilator. These factors must be taken into account before clinical use.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: