Ten different hip resurfacing systems: biomechanical analysis of design and material properties
Christian Heisel1, Jennifer A Kleinhans, Michael Menge
1Laboratory of Biomechanics and Implant Research, Department of Orthopaedics, University of Heidelberg, Heidelberg, Germany. christian.heisel@ok.uni-heidelberg.de
Abstract:
This study gives an overview of the main macro- and microstructural differences of ten commercially available total hip resurfacing implants. The heads and cups of resurfacing hip implants from ten different manufacturers were analysed. The components were measured in a coordinate measuring machine. The microstructure of the heads and cups was inspected by scanning electron microscopy. The mean radial clearance was 84.86 microm (range: 49.47-120.93 microm). The implants were classified into three groups (low, medium and high clearance). All implants showed a deviation of roundness of less than 10 microm. It was shown that all implants differ from each other and a final conclusion about the ideal design and material combination cannot be given based on biomechanical data. Widespread use of specific designs can only be recommended if clinical long-term follow-up studies are performed and analysed for each design.