Related Experiment Video
Updated: Jul 3, 2026

Optimized Griess Reaction for UV-Vis and Naked-eye Determination of Anti-malarial Primaquine
Published on: October 11, 2019
Simultaneous determination of artemether and lumefantrine in fixed dose combination tablets by HPLC with UV detection
Isabela da Costa César1, Fernando Henrique Andrade Nogueira, Gérson Antônio Pianetti
1Departamento de Produtos Farmacêuticos, Faculdade de Farmácia, Universidade Federal de Minas Gerais, Av. Pres. Antônio Carlos 6627, 31270-901 Belo Horizonte, MG, Brazil. isaccesar@bol.com.br
Abstract:
This paper describes the development and validation of a HPLC-UV method (210 nm) for the simultaneous quantitation of artemether and lumefantrine in fixed dose combination tablets. The method showed to be linear (r(2)>0.99), precise (R.S.D.<2.0%), accurate (recovery of 101.07% for artemether and 101.58% for lumefantrine), specific and robust. Four batches of artemether-lumefantrine tablets were assayed by the validated method. The artemether contents in the tablets varied from 98.61% to 103.35%, while lumefantrine contents were 97.92-100.48%.