Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
Accuracy of three-dimensional (3D) craniofacial cephalometric landmarks on a low-dose 3D computed tomograph
R Olszewski1, H Reychler, G Cosnard
1Department of Oral and Maxillofacial Surgery , Av. Hippocrate 10, Saint Luc University Clinics, Université Catholique de Louvain, 1200 Brussels, Belgium. raphael.olszewski@stom.ucl.ac.be
Objectives:
The aim of this paper is to compare the accuracy of cephalometric landmark identification using three-dimensional CT (3D-CT) surface rendering with "high-dose" (200 mAs) and "low-dose" (35 mAs) CT protocols. The absorbed dose levels for radiosensitive organs in the maxillofacial region during the exposure for both 3D-CT protocols were also measured.
Methods:
The study population consisted of 15 human dry skulls examined with spiral 3D-CT. 12 cephalometric anatomical landmarks at 7 sites were identified on the 3D-CT surface renderings by 2 observers independently, twice each, using high-dose and low-dose protocols. In total, 1440 imaging measurements were made. Thermoluminescent dosemeters (TLDs) were placed at ten sites around the thyroid and submandibular glands and the eyes in an Alderson phantom for measuring the absorbed dose levels.
Results:
The intraobserver mean distances between 3D landmarks were smaller for all sites with the high-dose protocol (P = 0.37). There was a significant difference among the observers (P = 0.000004). Interobserver mean distances between 3D landmarks were smaller for four of the seven sites with the low-dose protocol. However, the global interobserver mean distances between 3D landmarks for all sites were smaller with the high-dose protocol (P = 0.028). The low-dose protocol reduced the radiation dose to the thyroid by 6.12, to the submandibular salivary glands by 5.91 and to the eye by 5.44, resulting in a global reduction factor of 5.71.
Conclusions:
The accuracy in the landmark's identification was maintained when the milliampere-second values were reduced from 200 mAs to 35 mAs. We recommend use of the low-dose protocol for clinical 3D-CT cephalometric applications.
More Related Videos
08:03Midface Hypoplasia and Cranial Base Morphology in Syndromic Craniosynostosis: A Comparative Analysis Study Using a Predictive Regression Model
Published on: November 4, 2025
05:49Reliability of Artificial Intelligence-Based Cone Beam Computed Tomography Integration with Digital Dental Images
Published on: February 23, 2024
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...