Dosimetry of the cone beam computed tomography Veraviewepocs 3D compared with the 3D Accuitomo in different fields of
1Department of Dentomaxillofacial Radiology, Dental School, University of Leipzig, Germany.
Objectives:
This study compares tissue-absorbed and effective doses of the cone beam CT (CBCT) units, the Veraviewepocs 3D and the 3D Accuitomo, in different protocols.
Methods:
The absorbed organ doses were measured using an anthropomorphic phantom loaded with thermoluminescent dosemeters (TLDs) in 16 sensitive organ sites. Both CBCT units were deployed with different fields of view (FOVs): 3D Accuitomo using two protocols (anterior 4 x 4 cm scan and anterior 6 x 6 cm scan) and Veraviewepocs 3D using three protocols (anterior 4 x 4 cm scan, anterior 8 x 4 cm scan and panoramic + anterior 4 x 4 cm). Equivalent and effective doses were then calculated, the latter based on the International Commission on Radiological Protection's (ICRP) 2005 recommendations.
Results:
The lowest effective dose was observed for the 3D Accuitomo 4 x 4 cm (20.02 microSv), the highest for the 3D Accuitomo 6 x 6 cm (43.27 microSv). The effective dose recorded for Veraviewepocs 3D was 39.92 microSv for the 8 x 4 cm scan, 30.92 microSv for the 4 x 4 cm scan and 29.78 microSv for the panoramic + 4 x 4 cm scan protocol.
Conclusions:
The radiation doses delivered by both machines were in comparable ranges when using 4 x 4 cm FOV. A smaller FOV should be used for dental images, whereas a larger FOV should be restricted to cases in which a wider view is required.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...