Related Experiment Video
Updated: Jul 3, 2026

The Flexible Rhino-Laryngoscope for Awake Nasotracheal Intubation
Published on: August 2, 2024
Evaluation of the size 4 i-gel airway in one hundred non-paralysed patients
J J Gatward1, T M Cook, C Seller
1Department of Anaesthesia, Royal United Hospital, Combe Park, Bath, UK.
Abstract:
We studied the i-gel in 100 elective, anaesthetised patients (55:45 male:female, median age 53 years) assessing: ease of use, airway quality, positioning, seal and complications. First insertion attempt was successful in 86 patients, second attempt in 11 patients, and third attempt in three patients. Fifty three manipulations were required in 26 patients (median 1) to achieve a clear airway. Median insertion time was 15 s. During ventilation, expired tidal volume of 7 ml x kg(-1) was achieved in 96 (96%) patients. Median airway leak pressure was 24 cmH(2)O. On fibreoptic examination via the device, vocal cords were visible in 87 patients (91%). During maintenance, six patients (6%) required 12 airway manipulations. There was one episode of regurgitation, without aspiration. Other complications and patient side-effects were mild and few. The i-gel is easily and rapidly inserted, providing a reliable airway in over 90% of cases. Further studies are indicated to assess safety and performance compared to other supraglottic airway devices.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...
Aneurysm IV: Nursing Management
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Suctioning the Nasopharyngeal Airway
Equipment Required