T Bach1, B Geavlete, T R W Herrmann
1Asklepios Hospital Barmbek, Department of Urology, Hamburg, Germany. t.bach@asklepios.com
Tool selection significantly impacts flexible ureteroscope performance. Larger tools reduce irrigation flow and deflection, potentially compromising visualization during stone procedures.
Area of Science:
- Urology
- Medical Devices
- Endoscopy
Background:
- Flexible renoscopes provide access to the renal collecting system.
- Optimal visualization is crucial for stone workup, making irrigation flow critical.
Purpose of the Study:
- To evaluate the impact of tools and probes on deflection angle and irrigation flow volume in modern flexible renoscopes.
Main Methods:
- Tested five flexible renoscopes (Wolf, Storz, ACMI, Olympus).
- Measured deflection angle, bent tip diameter, and flow rates with empty and loaded scopes.
- Evaluated laser fibers, biopsy forceps, and nitinol baskets of various French sizes.
Main Results:
- Deflection angle is not affected by basket size but is reduced by laser fibers and biopsy forceps.
- Tool introduction increases bent tip diameter and significantly decreases irrigation flow volume, especially with larger tools.
- Irrigation volume loss ranged from 62% to 99% depending on tool size.
Conclusions:
- Intraoperative visualization is dependent on the working tool used due to effects on irrigation flow and deflection.
- Proper selection of flexible renoscopes, probes, and tools is essential for successful ureterorenoscopy.