Transmural surgical gown pressure measurements in the operating theater

K W Altman1, J H McElhaney, J A Moylan

  • 1Department of Biomedical Engineering, Duke University, Durham, NC.

Insights

Surgical gown pressure testing may be inaccurate. Peak intraoperative pressures exceed 60 psi, far surpassing the 5 psi industry standard, potentially compromising bacterial barrier efficacy.

Area of Science:

  • Surgical sciences
  • Biomedical engineering
  • Infection control

Background:

  • Surgical gowns and drapes are critical barriers against bacterial transmission.
  • Current industry standards for testing barrier efficacy assume low contact pressures (<5 psi).

Purpose of the Study:

  • To measure actual transmural gown pressures during surgical procedures.
  • To evaluate the adequacy of current industry testing standards for surgical gowns and drapes.

Main Methods:

  • Utilized pressure-sensitive contact film and resistive strain gauges.
  • Recorded pressures from the surgeon's abdominal region and forearms during operations.

Main Results:

  • Peak contact pressures recorded exceeded 60 psi.
  • Measured pressures were significantly higher than the assumed 5 psi industry standard.

Conclusions:

  • Current testing methods for surgical gown and drape barrier materials may not accurately reflect real-world conditions.
  • Reassessment of test utilization is needed to ensure effective bacterial barrier performance.