The performance of six pulse oximeters in the environment of neuronavigation
Alexander M Mathes1, Sascha Kreuer, Sven O Schneider
1Department of Anesthesiology, Critical Care and Pain Medicine, Saarland University Hospital, Homburg (Saar), Germany.
Background:
Although the use of pulse oximeters may be regarded a standard of care for monitoring anesthesia procedures, these monitors may be susceptible to various kinds of disturbances. Recently, it was suggested that neuronavigation equipment may interfere with pulse oximeter accuracy. In this study, we evaluated the effect of a neurosurgical image guidance system on the performance of six different pulse oximeters. Two simple shielding methods were evaluated.
Methods:
Twenty healthy, adult, nonsmoking volunteers were equipped with six different pulse oximeters on both hands. Baseline values for heart rate, arterial oxygen saturation, and signal quality were assessed. After activation of the Brain Lab VectorVision Neuronavigation System, the effects on signal quality and saturation recognition were evaluated. Measurements were repeated using two different shielding techniques, a cotton blanket and aluminum sheets.
Results:
Activation of the image guidance system resulted in a significant disturbance of signal quality and saturation detection, which was partially reversible by both shielding techniques. Significant differences were noted among the six brands of pulse oximeters for signal quality (P < 0.001) and saturation recognition (P < 0.001), and for the response to shielding methods (P < 0.001). Coverage of the probes with aluminum foil resulted an in undisturbed saturation recognition in all subjects with almost all monitors.
Conclusions:
Infrared pulse waves from neurosurgical navigation equipment may interfere with pulse oximeter measurements. Shielding the probe with aluminum foil sufficiently eliminated the infrared interference.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...