An evaluation of a new combined Spo2/PtcCO2 sensor in very low birth weight infants
Serafina Lacerenza1, Maria Pia De Carolis, Francesca Paola Fusco
1Division of Neonatology, Department of Pediatrics, Institute of Hygiene, Catholic University of Sacred Heart, Rome, Italy.
Background:
Recently, a new sensor for combined assessment of pulse oximetry oxygen saturation (Spo(2)) and transcutaneous monitoring of carbon dioxide partial pressure (PtcCO(2)) has been introduced (TOSCA 500, Radiometer basel AG, Switzerland) [corrected] We designed this study to evaluate the usability and reliability of TOSCA in neonates with birth weight
Methods:
In a prospective study of 22 newborns, TOSCA was tested, positioning the sensor on the ear pinna with an adhesive attachment clip. Simultaneous monitoring with TOSCA, conventional pulse oximeter (HP; Datex Ohmeda 3740), and a transcutaneous device (TINA TCM3, Radiometer, Copenhagen) was performed for 60 min. PtcCO(2) measurement from TOSCA (PtcCO(2TOSCA)) and TINA (PtcCO(2)) were compared with Pco(2) from blood samples (PCO(2EAB)) at 1 and 60 min. During the monitoring period, values of PtcCO(2TOSCA) were compared with PtcCO(2), and SatO(2) values from TOSCA with those from a pulse oximeter. Corresponding data were compared using Bland-Altman analysis.
Results:
Bias (precision) at 1 min and at 60 min between PCO(2EAB) and PtcCO(2) values were 3.5 (12.4) mm Hg and 2.8 (10.2), respectively, whereas between PCO(2EAB) and PtcCO(2TOSCA) values were 18.3 (30.4) mm Hg and 1.8 (25) mm Hg. Bland-Altman analysis shows a better correspondence PtcCO(2)/PtcCO(2TOSCA) between 7 and 15 min. No significant differences were found between Spo(2) and SpO(2TOSCA).
Conclusions:
The TOSCA monitor is safe and easy to apply in very low birth weight newborns. The pulse oximeter measurements may be useful for titrating oxygen therapy. Pco(2) measurement with TOSCA is most useful as a trend and independent confirmation of arterial Pco(2) is required if an accurate value is needed.
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like asthma or COPD,...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: