Assessment of the validation of blood pressure monitors: a statistical reappraisal
Bruce A Friedman1, Bruce S Alpert, David Osborn
1GE Healthcare, Tampa, FL, USA. bruce.friedman@med.ge.com
Abstract:
Protocols to determine the accuracy of noninvasive blood pressure monitors have been published by the Association for the Advancement of Medical Instrumentation, the British Hypertension Society, and the European Society of Hypertension (ESH). We compared the statistical basis for each of these protocols. Although the Association for the Advancement of Medical Instrumentation and British Hypertension Society protocols require a minimum of 85 participants, the ESH protocol requires only 33 participants. The reduced sample size results in a reduction in statistical power from 98 to 70%, which brings into question the applicability of the ESH (International) protocol.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement