Related Experiment Video
Updated: Jul 3, 2026

Low Molecular Weight Protein Enrichment on Mesoporous Silica Thin Films for Biomarker Discovery
Published on: April 17, 2012
Sieve-based device for MALDI sample preparation. I. Influence of sample deposition conditions in oligonucleotide
Laura Molin1, Simone Cristoni, Sara Crotti
1ISB-Ion Source and Biotechnologies, Via Fantoli 16/15, 20138 Milano, Italy.
Abstract:
Spraying of oligonucleotide-matrix solutions through a stainless steel (ss) sieve (38 microm, 450 mesh) leads to the formation, on the matrix-assisted laser desorption/ionization (MALDI) sample holder, of uniformly distributed microcrystals, well separated from each other. When the resulting sample holder surface is irradiated by laser, abundant molecular species form, with a clear increase in both intensity and resolution with respect to values obtained by 'Dried Droplet', 'Double Layer', and 'Sandwich' deposition methods. In addition, unlike the usual situation, the sample is perfectly homogeneous, and identical spectra are obtained by irradiating different areas. On one hand, the data indicate that this method is highly effective for oligonucleotide MALDI analysis, and on the other, that it can be validly employed for fully automated MALDI procedures.
Related Concept Videos
MALDI-TOF Mass Spectrometry