Related Experiment Video
Updated: Jul 3, 2026

In Vitro Application of a Wireless Sensor in Flexion-Extension Gap Balance of Unicompartmental Knee Arthroplasty
Published on: May 5, 2023
The effect of bearing congruency, thickness and alignment on the stresses in unicompartmental knee replacements
D J Simpson1, H Gray, D D'Lima
1Nuffield Department of Orthopaedic Surgery, University of Oxford, Nuffield Orthopaedic Centre, Windmill Road, Oxford OX3 7LD, UK.
Background:
Unicompartmental knee replacement offers an effective treatment for patients with single compartment knee disease and is becoming an increasingly popular alternative to total knee replacement. An important cause of failure in a unicompartmental knee replacement implant is polyethylene wear. Significant contributory factors to the amount of polyethylene wear are contact stress, bearing alignment, congruency and thickness.
Methods:
Four different unicompartmental knee replacement implant designs (Fully-Congruent; Partially-Congruent; Non-Congruent-metal-backed; Non-Congruent-all-polyethylene) were inserted into a validated finite element model of a proximal tibia. The effect that bearing congruency, alignment and thickness had on the polyethylene stresses during a simulated step-up activity for each design was investigated. Additionally, contact pressures were compared to those calculated from Hertz elastic theory.
Findings:
Only the Fully-Congruent bearing experienced peak von Mises and contact stresses below the lower fatigue limit for polyethylene during the step-up activity. The highest polyethylene contact stresses were observed for the Partially-Congruent and Non-Congruent-metal-backed designs, which experienced approximately three times the polyethylene lower fatigue limit. Increasing the bearing thickness from 3.5mm to 8.5mm of the Non-Congruent design decreased the contact stresses in the bearing; however they did not fall below the lower fatigue limit for polyethylene. Good agreement between finite element and Hertz contact pressures was found.
Interpretation:
Fully congruent unicompartmental knee replacement bearings can be markedly thinner without approaching the material failure limit, have a greater potential to preserve bone stock and are less likely to fail mechanically.
Related Concept Videos
Bearing Stress
Due to the intricacy of these microforces, an average value, known as bearing stress, is often used by...
Unsymmetric Loading of Thin-Walled Members
The concept of the shear center is crucial in countering the...
Residual Stresses in Circular Shafts
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Bending of Members Made of Several Materials
Hooke's Law determines stress in each material, stating that stress is proportional to strain but varies due to each material's...