Related Experiment Video
Updated: Jul 3, 2026

Tuning a Parallel Segmented Flow Column and Enabling Multiplexed Detection
Published on: December 15, 2015
HPLC-UV method development and validation for the determination of low level formaldehyde in a drug substance
1Analytical Sciences Department, Research and Development Center, Boehringer Ingelheim Pharmaceuticals, Inc., Ridgefield, CT 06877, USA.
Abstract:
A reversed-phase high-performance liquid chromatographic method (HPLC) with diode-array detection is developed and validated for the quantitative determination of formaldehyde in a drug substance. Formaldehyde (HCHO) is reacted with 2,4-dinitrophenylhydrazine (DNPH) to form a Schiff base (HCHO-DNPH derivatization product), which has an absorbing maximum (lambda max) at 360 nm. The HPLC method employs a C8, 3-microm particle size analytical column (150 mm x 4.6 mm), 15-microL injection volume, column temperature controlled at 30 degrees C, detection at 360 nm, and a water-acetonitrile (55:45, v/v) mobile phase at a flow rate of 1 mL/min. These conditions resolve the HCHO-DNPH product from unreacted DNPH, the drug substance and related impurities, as well as diluent peaks within 20 min. The retention time of the HCHO-DNPH product is approximately 6.4 min. The method is linear, accurate in the specified range (0.33-333 ppm), and robust based on analyte (HCHO-DNPH derivatization product) stability in standard and sample. Detection limit is 0.03 ng (0.1 ppm).
More Related Videos
Related Concept Videos
Clinically Relevant Drug Product Specifications: Methods of Establishment
Data Validation
Key parameters for method validation include:
High-Performance Liquid Chromatography: Instrumentation