Related Experiment Video
Updated: Jul 3, 2026

Fabrication of Small Caliber Stent-grafts Using Electrospinning and Balloon Expandable Bare Metal Stents
Published on: October 26, 2016
Finite element comparison of performance related characteristics of balloon expandable stents
E W Donnelly1, M S Bruzzi, T Connolley
1Department of Mechanical & Biomedical Engineering, National University of Ireland, Galway, Ireland. evin.donnelly@nuigalway.ie
Abstract:
Cardiovascular stents are commonly made from 316L stainless steel and are deployed within stenosed arterial lesions using balloon expansion. Deployment involves inflating the balloon and plastically deforming the stent until the required diameter is obtained. This plastic deformation induces static stresses in the stent, which will remain for the lifetime of the device. In order to determine these stresses, finite element models of the unit cells of geometrically different, commercially available balloon expandable stents have been created, and deployment and elastic recoil have been simulated. In this work the residual stresses associated with deployment and recoil are compared for the various stent geometries, with a view to establishing appropriate initial stress states for fatigue loading for the stents. The maximum, minimum, and mean stresses induced in the stent due to systolic/diastolic pressure are evaluated, as are performance measures such as radial and longitudinal recoil.