Related Experiment Video
Updated: Jul 3, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Irregular ventricular tachycardia underdetected by implantable cardioverter defibrillator device
Athanasios G Manolis1, Spyros Kourouklis, Dimitrios Chatzis
12nd Department of Cardiology. Hellenic Red Cross General Hospital, Athens, Greece.
Abstract:
A case of sustained monomorphic ventricular tachycardia underdetected by a single chamber implantable cardioverter defibrillator because of RR interval irregularity is presented. The programmed stability criterion is responsible for the underdetection. Special attention must be paid when it comes to programming this detection parameter.
Related Concept Videos
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias III: Characteristics of Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Mechanism of Cardiac Arrhythmias
Dysrhythmias II: Classification of Tachyarrhythmias
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...