Cervical spine imaging using standard C-arm fluoroscopy: patient and surgeon exposure to ionizing radiation
Brian D Giordano1, Judith F Baumhauer, Thomas L Morgan
1Department of Orthopedics and Rehabilitation, University of Rochester Medical Center, Rochester, NY, USA.
Study Design:
A cadaveric cervical spine specimen is imaged with a standard C-arm fluoroscope during a simulated procedure. Patient and surgeon exposure to radiation is estimated by placing dosimeters at various locations in 3-dimensional space.
Objective:
The purpose of this study was to evaluate radiation exposure to patient and surgeon when using C-arm fluoroscopy during a simulated cadaveric surgical procedure involving the cervical spine.
Summary Of The Background Data:
The use of mobile fluoroscopy has become commonplace in orthopaedics. With the current trend towards minimal access techniques, fluoroscopy has become requisite to achieving satisfactory outcomes. Studies have shown that spine surgeons may be at elevated risk for radiation exposure compared to other orthopaedists. Exposure while using C-arm fluoroscopy for procedures involving the pelvis, as well as thoracic and lumbar spine has been documented. However, there are no equivalent studies that evaluate exposure during cervical spine imaging.
Methods:
A standard OEC 9800 C-arm was used to image a prepared cadaveric cervical spine specimen, which was suspended on an adjustable platform. Film badge dosimeters were mounted at various positions and angles to detect direct and scatter radiation. Testing was conducted in various radiation dose mapping "scenarios." The configurations tested altered the proximity of the specimen and jig relative to the radiation source. We attempted to capture radiation exposure in various locations, from a best-case to a worst-case scenario, as may be realistically encountered in a procedural setting. RESULTS.: Potential exposure to the patient and surgeon were consistently measurable, and of concern. As the imaged specimen was positioned closer to the radiation source, exposure to the patient was markedly amplified. Exposure to the surgeon did not increase as dramatically. There was a great degree of variability in the exposure doses recorded by the peripheral dosimeters. Even dosimeters that were placed in the same plane diverged widely in their measured exposure. This highlights the influence of the shape of the imaged specimen on reflected scatter. Scatter radiation doses on both sides of the specimen were similar.
Conclusion:
Care should be taken when working on both sides of the imaged subject. Considerable radiation exposure can be encountered when working with a C-arm fluoroscope if appropriate precautions are not observed. All appropriate radiation dose-reducing measures should be strictly enforced by the supervising physician to minimize risk to the patient and the medical team.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
X-ray Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...