Tono-Pen XL tonometry during application of a suction ring in rabbits
Spyridon K Charisis1, Harilaos S Ginis, Georgios A Kounis
1University Hospital of Heraklion, Eye Clinic, Heraklion, Crete, Greece. scharisis@gmail.com
Background:
The purpose of this study is to evaluate the use of Tono-Pen XL in measuring IOP during the application of a suction ring in rabbit eyes with manometrically controlled IOP.
Methods:
Tono-Pen XL was calibrated against direct manometry in 10 rabbit eyes. A suction ring was then applied in 4 rabbit eyes and the IOP was determined manometrically during suction ring application at 350 mmHg vacuum pressure. Finally, in 6 catheterized rabbit eyes the IOP was measured with Tono-Pen XL during suction ring application at suction vacuum from 350 to 650 mmHg, while keeping actual IOP stable at 30 mmHg and 60 mmHg.
Results:
Linear regression analysis revealed that the Tono-pen XL was reliable for IOPs between 10 and 70 mmHg (R2 = 0.9855). Direct manometry during suction ring application showed no statistically significant variation of Tono-Pen XL readings when the incanulation manometry intraocular pressure changed from 30 mmHg to 60 mmHg and no statistically significant correlation between suction vacuum and IOP measurements.
Conclusion:
Tono-Pen XL measurements are unreliable during the application of a suction ring on living rabbit eyes even when the actual IOP is forced to be within the validated range of Tono-Pen XL measurements. This inaccuracy is probably related to altered corneal and scleral geometry and stress.