Mechanical complications associated with mentor inflatable penile prosthesis.

W B Steinkohl1, G E Leach

  • 1Department of Urology, Kaiser Permanente Medical Center, Los Angeles, California.

Urology
|July 1, 1991
PubMed
Summary

The Mentor inflatable penile prosthesis shows improved mechanical reliability after a 1985 design modification. Post-modification, the failure rate dropped significantly, making the device more dependable for patients.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Vaginal repair of large cystoceles.

Current urology reports·2002
Same author

Undiagnosed urethral carcinoma: an unusual cause of female urinary retention.

Urology·2001
Same author

A new technique for cystocele repair and transvaginal sling: the cadaveric prolapse repair and sling (CAPS).

Urology·2000
Same author

Pelvic prolapse.

The Journal of urology·2000
Same author

Erosion of woven polyester pubovaginal sling.

The Journal of urology·1999
Same author

Stress urinary incontinence.

Current opinion in urology·1999

Area of Science:

  • Urology
  • Biomedical Engineering
  • Medical Device Reliability

Background:

  • Mechanical failures in penile prostheses can impact patient outcomes.
  • The Mentor inflatable penile prosthesis was designed to address these reliability issues.

Purpose of the Study:

  • To evaluate the mechanical reliability of the Mentor inflatable penile prosthesis.
  • To assess the impact of a 1985 design modification on device failure rates.

Main Methods:

  • Retrospective review of 46 patients implanted with the Mentor device between December 1982 and June 1987.
  • Analysis of mechanical failure rates before and after the April 1985 device modification.

Main Results:

  • Prior to the 1985 modification, 10 out of 30 (33.3%) implantations experienced mechanical failure, primarily due to input tubing cracks.

Related Experiment Videos

  • After the April 1985 design improvement, the mechanical failure rate decreased to 4%.
  • Conclusions:

    • The Mentor inflatable penile prosthesis demonstrates significantly improved mechanical reliability following its 1985 design enhancement.
    • The device modification effectively addressed the primary cause of mechanical failure, enhancing its long-term dependability.