Related Experiment Video
Updated: Jul 3, 2026

The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
Accuracy of MRI vs CT imaging with particular reference to patient specific templates for total knee replacement
D White1, K L Chelule, B B Seedhom
1Division of Bioengineering, Academic Unit of Musculoskeletal Diseases, Leeds Medical School, University of Leeds, UK. d.m.white@leeds.ac.uk
Background:
The present study compares the accuracy of MRI and CT imaging for the manufacture of patient-specific templates for total knee replacement surgery.
Methods:
A total of 10 ovine knees were imaged using MRI and CT scanners. Each set of images was reconstructed in 3D and then used to manufacture physical models of each bone, using rapid prototyping technology. After imaging the soft tissues were removed and specific measurements of the bony anatomy compared with measurements from the MRI and CT models.
Results:
Bone models generated from MRI scans were dimensionally less accurate than those generated from CT scans. Furthermore, the bone models generated from MRI scans were visibly inferior to those generated from the CT scans.
Conclusions:
Current MRI scans do not offer a viable alternative to CT. Although adaptation of the template system to accommodate MRI imaging is possible, the changes required are neither practical nor desirable.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Magnetic Resonance Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...