Left ventricular ejection fraction: single-plane and multiplanar transesophageal echocardiography versus equilibrium
M L Nessly1, G Bashein, P R Detmer
1Department of Anesthesiology, University of Washington School of Medicine, Seattle.
Abstract:
The relative accuracy and precision of estimating left ventricular ejection fraction (EF) in dogs were assessed by two-dimensional transesophageal echocardiography (2D-TEE) and by three-dimensional transesophageal echocardiographic (3D-TEE) imaging and reconstruction. This assessment was accomplished by comparing each echocardiographic method to a gated equilibrium blood pool radionuclide (RN) standard. By using both correlation and regression analysis, 2D-TEE performed reasonably well in estimating RNEF (correlation coefficient [r] = 0.80, slope = 1.01, intercept = 6.37, standard error of the estimate [SEE], 8.98), but not as well as 3D-TEE (r = 0.86, slope = 0.83, intercept = 3.38, SEE, 5.74). Using Altman and Bland's methods of comparison analysis, it was found that 2D-TEE overestimated RNEF by 7% (standard deviation [SD], 8.8). This degree of overestimation was not consistent across the range of measurement. In contrast, 3D-TEE slightly underestimated RNEF by less than 3% and showed less variability (SD, 6.0). The accuracy of the 3D-TEE determinations was not dependent on the magnitude of EF. Additionally, a significantly higher proportion of the 2D-TEE measurements (0.30) compared with the 3D-TEE measurements (0.10) differed from RN values by more than 10% (P = 0.009, McNemar's test). At the clinically important low end of the EF range (RNEF less than or equal to 35%), 2D-TEE may be expected (with 95% confidence) to be within -15% to +28% EF of reference values, whereas 3D-TEE can be expected to be within -8% to +5% EF relative to RN.(ABSTRACT TRUNCATED AT 250 WORDS)