Related Experiment Video
Updated: Jul 3, 2026

Remote Magnetic Navigation for Accurate, Real-time Catheter Positioning and Ablation in Cardiac Electrophysiology Procedures
Published on: April 21, 2013
Remote magnetic navigation-assisted catheter ablation enhances catheter stability and ablation success with lower
Darryl R Davis1, Anthony S L Tang, Michael H Gollob
1Department of Cardiology, University of Ottawa Heart Institute, Ottawa, Ontario, Canada.
Background:
It has been suggested that remote magnetic navigation (RMN) may provide enhanced catheter stability and substrate contact to aid in ablation. To date, no study has examined this claim. Accordingly, we compared the characteristics of the successful ablation of atrioventricular reentry tachycardia (AVNRT) using RMN with a matched population ablated using a conventional (CON) manual approach.
Methods:
Sixteen patients who underwent RMN-assisted ablation of typical AVNRT were matched with 16 patients who had a CON-AVNRT ablation.
Results:
All patients had successful slow pathway modification without complication. The mean catheter temperature achieved with the successful ablation was significantly lower with RMN than with CON (42 +/- 7 degrees C vs 47 +/- 3 degrees C, P
Conclusions:
Although the construction of the ablation catheters is similar, ablations with RMN catheters resulted in a lower mean temperature, earlier time to JT, and less variability of temperature during ablation, suggesting greater catheter stability. This study indicates that ablation with RMN can achieve success with lower catheter temperatures.