Poor mid-term results of total hip arthroplasty with use of a Hylamer liner
Related Experiment Video
Updated: Jul 3, 2026

Surgical Technique of the 3-Dimensional-printed Personalized Hip Implant for the Treatment of Canine Hip Dysplasia
Published on: April 19, 2024
Adrian Skwara1, Simone Stracke, Carsten O Tibesku
1Department of Orthopaedics and Rheumatology, University Hospital Marburg, Germany. skwara@med.uni-marburg.de
Abstract:
Clinical results after total hip arthroplasty with the use of a Hylamer liner have been described as being poor. In this study, polyethylene wear rates as well as clinical results were assessed after hip arthroplasty using a Hylamer liner. Thirty patients were followed radiographically and clinically after total hip arthroplasty with a Hylamer liner. The mean follow-up was six years. Radiolucent lines and wear rates were evaluated in correlation to different parameters and clinical scores. Radiolucent zones were found in 40% of the femoral and 27% of the acetabular components. Wear rates showed no correlation to the clinical scores, nor to other parameters. The SF-36 questionnaire showed inferior results compared to healthy controls. These findings confirm the poor results described with Hylamer in current literature. A significant influence of the different patient or prosthesis specific parameters on the survival of the prosthesis could not be confirmed.