Evaluation of Keeler Pulsair non-contact tonometry: reliability and reproducibility
L Bonomi1, S Baravelli, C Cobbe
1Institute of Ophthalmology, University of Verona, Italy.
Abstract:
Intraocular pressure was evaluated in 414 eyes using Keeler Pulsair tonometry in comparison with values obtained by Goldmann tonometry. No statistically significant differences were found between the two techniques. Landis K statistical analysis was used to evaluate the measurements obtained by two operators in 125 eyes, with the results indicating good reproducibility. In 20 eyes, a statistically significant difference was found between values obtained with patients in the seated vs supine positions, whereas measurements made with subjects' heads in a supported or unsupported position yielded no statistically different values.