Evaluation of Fluid Overload by Bioelectrical Impedance Vectorial Analysis
Published on: August 17, 2022
Subcutaneous bioimpedance recording: assessment of a method for hemodynamic monitoring by implanted devices
Deviprasad Venugopal1, Robert Patterson, Rajat Jhanjee
1Departments of Medicine (Cardiac Arrhythmia Center), University of Minnesota Medical School, Minneapolis, Minnesota 55455, USA.
Background:
Diagnostic evaluation of patients with suspected symptomatic arrhythmias is limited by inability to assess the hemodynamic impact of a detected rhythm.
Objective:
To address this limitation, we utilized closely spaced subcutaneous electrodes, small enough to incorporate within an implantable monitor, to detect blood flow-induced pectoral muscle bioimpedance (Z) changes in a swine model of hemorrhage-induced hypotension.
Methods:
In seven anesthetized and ventilated adult pigs, small ring electrodes (current electrodes 5 cm apart; voltage electrodes 3.5 cm apart) were positioned on the left pectoral muscle. Z signals (Biopac system) and invasive arterial blood pressures were recorded. Hypotension was induced by hemorrhage (50% blood volume reduction). Mean arterial pressure (MAP) and pulse pressure (PP) with corresponding pulse Z (DeltaZ) and base Z (Z(o)) were measured. A longitudinal mixed model with a first-order autoregressive error structure was used to test for associations (change in DeltaZ vs change in MAP and change in DeltaZ vs change in PP) taking into account within pig correlation.
Results:
During bleeding-induced hypotension, Z(o) increased. Changes of DeltaZ correlated with both a change in MAP (coefficient = 1.17, P < 0.0001) and change in PP (coefficient = 0.98, P < 0.0001). A change in DeltaZ of 1-2 orders of magnitude corresponded to an approximate 40-70% drop in MAP and PP in a porcine model in which the baseline MAP was 69-70 mmHg.
Conclusion:
Our findings suggest that closely spaced subcutaneous electrodes identify changes in local tissue/vascular bioimpedance that correlate well with direct invasive measures of induced hypotension in a porcine model.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...