The impact of scan length on the exposure levels in 16- and 64-row multidetector computed tomography: a phantom study
Roy Irwan1, Henk B K de Vries, Paul E Sijens
1Toshiba Medical Systems Europe, Zoetermeer, The Netherlands.
Rationale And Objectives:
Higher patient exposure levels have been reported for 64-row multidetector computed tomography (MDCT) compared to 16-row MDCT. The objective of this study was to make a thorough comparison by evaluating the impact of scan length on the exposure levels at 16-row MDCT and 64-row MDCT.
Materials And Methods:
Dose-length product (DLP) values were determined to compare exposure levels in 16- and 64-row MDCT. This phantom study does not deal with a possible reduction in image quality induced by an increase in scattered radiation in 64-row MDCT compared to 16-row MDCT.
Results:
The exposure levels of 64-row MDCT (scan slice thickness, 0.5 mm) are up to 18% lower than those of 16-row MDCT at slice thickness 0.5 mm when scanning an object larger than 12.3 cm. At this value, the plots of the 16- and 64-row DLP values versus scan length cross. The DLP curves of 1- and 2-mm slice thickness 16-row MDCT are in closer resemblance to those of 0.5-mm 64-row MDCT. The respective exposure levels of 1- and 2-mm slice thickness 16-row MDCT exceed those of 0.5-mm 64-row MDCT by up to 4% and 3%, with intersections of 30 and 25 cm, respectively.
Conclusion:
Lower effective doses are obtained in 64-row MDCT compared to 16-row MDCT (0.5-mm slice thickness) provided that scan length exceeds 12.3, 30, and 25 cm, for 16-row MDCT slice thickness of 0.5, 1, and 2 mm, respectively. Reduced effective dosage in 64-row MDCT compared to 16-row MDCT has not been demonstrated before. Differences in object size may thus explain discrepancies between previous studies with regard to the exposure levels at 64-slice CT compared to 16-slice CT.
More Related Videos
11:49Using High Resolution Computed Tomography to Visualize the Three Dimensional Structure and Function of Plant Vasculature
Published on: April 5, 2013
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT