Related Experiment Video
Updated: Jul 2, 2026

An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Bovine bone screws: metrology and effects of chemical processing and ethylene oxide sterilization on bone surface and
D P Haje1, J B Volpon, C A Moro
1Clinical Center Orthopectus, SMHN-Q.2 Bl. A, Ed. de Clínicas 804/806, 70710-904, Brasilia, DF, Brazil. davihaje@yahoo.com.br
Abstract:
We assess the effects of chemical processing, ethylene oxide sterilization, and threading on bone surface and mechanical properties of bovine undecalcified bone screws. In addition, we evaluate the possibility of manufacturing bone screws with predefined dimensions. Scanning electronic microscopic images show that chemical processing and ethylene oxide treatment causes collagen fiber amalgamation on the bone surface. Processed screws hold higher ultimate loads under bending and torsion than the in natura bone group, with no change in pull-out strength between groups. Threading significantly reduces deformation and bone strength under torsion. Metrological data demonstrate the possibility of manufacturing bone screws with standardized dimensions.