Electron multiplier-scintillator detector for pulse counting positive or negative ions
1General Electric Company, Knolls Atomic Power Laboratory, Schenectady, New York 12301.
The Review of Scientific Instruments
|September 1, 1978
Summary
A novel hybrid ion detector combines electron multipliers and scintillator detectors for enhanced mass spectrometry. This new design offers superior performance and low noise for fast pulse counting across a wide range of ion masses.
Area of Science:
- Analytical Chemistry
- Instrumentation Science
Background:
- Conventional ion detectors in mass spectrometry, such as electron multipliers and scintillator detectors, have limitations in performance and flexibility.
- There is a need for a versatile and high-performance ion detector capable of fast pulse counting for various mass spectrometry applications.
Purpose of the Study:
- To develop and characterize a novel general-purpose ion detector for fast pulse counting in mass spectrometry.
- To combine the advantages of electron multipliers and scintillator detectors into a single hybrid instrument.
Main Methods:
- Development of a hybrid detector integrating a six-stage electron multiplier with a scintillator detector.
- Implementation of post-acceleration of mass-resolved ions to 10 keV before detection.
- Testing of the detector's performance in both positive and negative ion detection modes.
Main Results:
- The hybrid detector demonstrates superior performance compared to conventional electron multiplier or scintillator detectors.
- Achieved background counting noise of less than 1 count/min.
- Efficient detection of ions from 1 amu to several hundred amu with near-unity efficiency for low-energy ions.
- Capability for counting randomly spaced pulses from 0 Hz to 10 MHz.
Conclusions:
- The developed hybrid ion detector offers a significant advancement in mass spectrometry instrumentation.
- Its unique design provides high sensitivity, low noise, and broad applicability for ion detection.
- This detector is suitable for a wide range of mass spectrometry applications requiring fast and accurate ion counting.
Related Concept Videos
Mass Analyzers: Common Types
The quadrupole mass analyzer consists of four cylindrical metal rods arranged in a diamond carrying a DC voltage and a radio-frequency AC voltage. The motion of ions through the quadrupole depends on the field strength, causing only ions of a certain m/z to resonate successfully and strike the detector at a given field strength. Though the transmission rate for these analyzers is high, the exact elemental composition of the sample is not determined because of low resolution; however, they are...
Mass Spectrometers
This lesson details the instrumentation of a mass spectrometer—a physical instrument to perform mass spectrometry on analyte molecules and record the characteristic mass spectra. This is achieved via three chief functions:
Gas Chromatography: Types of Detectors-II
In gas chromatography, different detectors are employed to meet specific analytical needs. These detectors are often categorized based on their detection mechanisms and the types of compounds they are best suited to analyze. Thermal Conductivity Detectors (TCD), Flame Ionization Detectors (FID), and Electron Capture Detectors (ECD) represent common categories, each with unique operating principles and applications. However, beyond these, several other detectors are designed for more specialized...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
In inductively coupled plasma–mass spectrometry (ICP–MS), an inductively coupled plasma (ICP) torch is used as an atomizer and ionizer. Solid samples are dissolved and volatilized before being introduced into the high-temperature argon plasma, while solution samples are nebulized and passed through the high-temperature argon plasma. Plasma dissociates the analytes and ionizes their component atoms to form a mixture of positive ions and molecular species. The positive ions are then passed on to...
Electrospray Ionization (ESI) Mass Spectrometry
Higher molecular weight biomolecules are nonvolatile compounds that may decompose before ionizing or vaporizing during mass analysis with conventional electron impact ionization methods. Accordingly, electrospray ionization (ESI) is the favored method for vaporizing and ionizing biomolecules as it circumvents rapid fragmentation and enables the recording of mass signals for the entire biomolecule.
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Tandem Mass Spectrometry
Tandem mass spectrometry is a technique that uses multiple mass analyzers in series to obtain a higher selectivity and reduce chemical noise during analyte detection. Instruments with multiple analyzers separated by an interaction cell enable secondary fragmentation and selected study of the fragment ions.Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called...