C M Brodbeck1, R R Bukrey, J T Hoeksema
1Department of Physics, Loyola University of Chicago, Chicago, Illinois 60626.
Abstract:
A versatile ac mutual inductance bridge is described which uses integrated circuit operational amplifiers (op-amps) for the measurement of magnetic susceptibilities of ferro-, ferri-, and para-magnetic samples. The circuit employs op-amps both for balancing the bridge and for detecting the differential signal from the sample coil. The response of the circuit is linear over a wide range of sample susceptibilities, and is calibrated directly in absolute units (emu) using a multiturn digital potentiometer. The sensitivity of the instrument for weak paramagnetic samples (1 g) is +/-3x10(-7) emu.
Related Concept Videos
Mutual Inductance
When two circuits carrying time-varying currents are close to one another, the magnetic flux through each circuit varies because of the changing current in the other circuit. Consequently, an emf is induced in each circuit by the changing current in the other. Therefore, this type of emf is called...
Inductor in an AC Circuit
Calculation of Self-inductance
Since the effect of the induced electric field and the back EMF generated depends on the rate of change of current and the self-inductance, the inductance...
Magnetic Field Of A Current Loop
Self-Inductance
Consider a circuit connected to an AC source. As the current varies with time, the magnetic flux through the circuit correspondingly changes. Faraday's law tells us that an emf would therefore...
Magnetic Susceptibility and Permeability
When diamagnetic materials are placed under an external magnetic field, the moments opposite to the field are induced. Hence, the susceptibility for diamagnets has a minimal negative value of 10-5–10-6. Since...