High resolution T2 weighted liver MR imaging using functional residual capacity breath-hold with a 1.0-Tesla scanner
Akihiko Tabuchi1, Toshizo Katsuda, Rumi Gotanda
1Graduate School of Health Sciences, Okayama University, Okayama, Japan. a-tab@po1.oninet.ne.jp
Purpose:
During acquisition of rapid high resolution (HR) T2 weighted (T2W) liver magnetic resonance (MR) images using a 1.0-Tesla (T) scanner, the liver is segmented into odd and even sections that are acquired at two different times using the multi-breath-hold (MBH) strategy. Misalignment between the two breath-hold (B-H) images may result in the occurrence of a blind area and a decrease in diagnostic accuracy. Here, a functional residual capacity (FRC) B-H method was developed to overcome this problem.
Material And Methods:
Twenty-five volunteers were enrolled. The sagittal images were reconstructed from whole liver transverse images. When the B-H phases are different, misalignment may occur in the craniocaudal and anterior-posterior (AP) directions. In this study, misalignments of the abdominal wall were measured in the AP direction. The misalignment was compared between four B-H phases, maximum inspiration (MI), maximum expiration (ME), voluntary expiration (VE) and FRC using one-way repeated measures ANOVA. Differences between groups were compared using the t-test for multi-group comparisons. In addition, qualitative analysis of misalignment was performed between VE and FRC in 52 clinical patients and the chi(2) test was performed.
Results:
The misalignment widths of FRC, ME, MI and VE B-Hs were 2.7+/-3.8, 6.4+/-7.4, 9.1+/-8.4 and 6.0+/-6.7 mm, respectively. Misalignment of the liver position using FRC was significantly smaller than for the other B-H methods (p<0.05). Significant differences between the VE B-H and FRC B-H were also observed in the qualitative analysis (p<0.05).
Conclusion:
The liver positions obtained when using FRC B-H were significantly more reproducible than when using the other B-H methods. The FRC B-H method resulted in a reduction in the blind area and an extension of the diagnostic area to the whole liver.