Identifying Coronary Artery Calcification on Non-gated Computed Tomography Scans
Published on: August 28, 2018
Calcified vascular plaque specimens: assessment with cardiac dual-energy multidetector CT in anthropomorphically
Daniel T Boll1, Elmar M Merkle, Erik K Paulson
1Department of Radiology, Duke University Medical Center, DUMC 3808, Durham, NC 27710, USA. daniel.boll@duke.edu
Purpose:
To evaluate whether dual-energy multidetector computed tomography (CT) with image postprocessing techniques enhances accuracy of calcified plaque quantification beyond the scope of single-energy multidetector CT, by using optical coherence tomography (OCT) as the reference standard.
Materials And Methods:
Four atherosclerotic specimens were examined with 64-section dual-energy multidetector CT by using a novel dual-detector "double-decker" design, with stacked high- and low-energy detector arrays with 32 x 0.625-mm collimation, at 140 kVp and 400 mAs, acquiring simultaneous and isopedic low- and high-energy data sets. Additionally, combined-energy data sets were calculated, and an enhancement algorithm was proposed. Cardiac motion was simulated by an anthropomorphically moving phantom, and OCT was used as a reference standard for plaque quantification. Univariate general linear model (GLM) analysis was used to compare sizes of plaque calcifications determined with OCT with those determined with dual-energy multidetector CT, and the significance of factors such as cardiac motion was assessed.
Results:
GLM analysis revealed that plaque quantification based on low-, high-, and combined-energy data sets differed significantly from that based on OCT (P < .001). Greater data variation occurred in smaller (<8 mm(2)) and larger (>12 mm(2)) calcifications. Comparison of calcified plaque sizes determined with OCT with those determined with the dual-energy multidetector CT enhancement algorithm revealed no significant difference (P = .550). Cardiac activity led to a slight increase in data variation in regard to OCT for corresponding static (mean, 10.2% +/- 3.2 [standard deviation]) and dynamic (13.8% +/- 4.9) dual-energy multidetector CT data sets.
Conclusion:
Dual-energy multidetector CT with novel postprocessing techniques enhanced accuracy of calcified plaque quantification by reducing effects of tissue blooming and beam hardening beyond single-energy multidetector CT.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System IV: CMRI