Related Experiment Video
Updated: Jul 2, 2026

A Rapid and Chemical-free Hemoglobin Assay with Photothermal Angular Light Scattering
Published on: December 7, 2016
Direct electrochemistry of hemoglobin on carbonized titania nanotubes and its application in a sensitive reagentless
Chunxian Guo1, Fengping Hu, Chang Ming Li
1School of Chemical and Biomedical Engineering, Nanyang Technological University, Singapore 637457, Singapore.
Abstract:
Carbonized TiO(2) nanotubes (TNT/C) prepared by carbonization with organic polymers possess advantages combined from high conductivity of carbon and nanostructure of TiO(2) nanotubes. The material was used as a supporting matrix to immobilize a redox protein, hemoglobin (Hb), to explore its direct electron transfer ability. The apparent heterogeneous electron transfer rate constant (k(ET)) of Hb on TNT/C is 108s(-1), which is much higher than that in the reported works, demonstrating excellent direct electrochemistry behavior. The TNT/C-Hb modified glassy carbon electrode (GCE) demonstrates significant electrocatalytic activity for reduction of hydrogen peroxide with a small apparent Michaelis-Menten constant (87.5 microM). The TNT/C-Hb based H(2)O(2) sensor has a low detection limit (0.92 microM), fast response time (3s) and high dynamic response range (10(-6) to 10(-4)M), a much better performance than the reported works. These results demonstrate that a direct electrochemistry behavior can be significantly enhanced through simple carbon coating on a nanostructured material for higher reaction surface area and better conductivity. This work suggests that Hb-immobilized TNT/C has potential applications in a sensitive H(2)O(2) sensor.