Dose related, comparative evaluation of a novel bone-subtraction algorithm in 64-row cervico-cranial CT angiography
1Department of Neuroradiology, Charité Universitary Medicine Berlin, Germany.
Purpose:
Comparative evaluation of a low-dose scan protocol for a novel bone-subtraction (BS) algorithm, applicable to 64-row cervico-cranial (cc) CT angiography (MSCTA).
Methods And Patients:
BS algorithm assessment was performed in cadaveric phantom studies by stepwise variation of tube current and head malrotation using a 64-row CT scanner. In order to define minimum dose requirements and the rotation correction capacity, a low dose BS MSCTA protocol was defined and evaluated in 12 patients in comparison to a common manual bone removal algorithm. Standard MIPs of both modalities were evaluated in a blinded manner by two neuroradiologists for image quality composed, of vessel contour sharpness and bony vessel superposition, by using a five-point score each. Effective Dose (E) and data post-processing times were defined.
Results:
In experimental studies prescan tube current could be cut down to one-sixth of post-contrast scan doses without compromise of bone-subtraction whereas incomplete subtraction appeared from four degrees head malrotation on. Prescan E amounted to additional 1.1mSv (+25%) in clinical studies. BS MSCTA performed significantly superior in terms of bony superposition for vascular segments C3-C7 (p<0.001), V1-V2, V3-V4 (p<0.05, p<0.001 respectively) and the ophthalmic artery (p<0.05), whereas vessel contour sharpness in BS MSCTA only proved superior for arterial segments V3-V4 (p<0.001) and C3-C7 (p<0.001). MBR MSCTA received higher ratings in vessel contour sharpness for C1-C2 (p<0.001), callosomarginal artery (p<0.001), M1, M2, M3 (p<0.001 each) and the basilar artery (p<0.001). Reconstruction times amounted to an average of 1.5 (BS MSCTA) and 3min (MBR MSCTA) respectively.
Conclusion:
The novel BS algorithm provides superior skull base artery visualisation as compared to common manual bone removal algorithms, increasing the Effective Dose by one-fourth. Yet, inferior vessel contour sharpness was noted intracranially, thus limiting the BS algorithm use to patients with suspected vessel pathology at the skull base level.
More Related Videos
05:49Reliability of Artificial Intelligence-Based Cone Beam Computed Tomography Integration with Digital Dental Images
Published on: February 23, 2024
08:39Longitudinal Micro-Computed Tomography Image Analysis for User-Defined Region of Interest in Critical-Sized Bone Defects
Published on: June 24, 2025
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT